![]() ![]() TB 9-4910-556-35
Table 1. Calibration Description - Continued
Test instrument parameters
Performance specifications
Controllable interface box - continued
100 mV channel
Range: 0 to 100 mV dc
Accuracy: ±5 MV
0 V stimulus
Range: 0 V dc
Accuracy: ±0.083 V
9.5 V stimulus
Range: 9.5 V dc
Accuracy: ±0.2 V
VTM voltage probe
Range: 24 V dc
Accuracy: ±1.2 V
Set communicator
Dc voltage
Range: 0.000 to 1.999 V
Accuracy: ±(1.0% of indication +0.2% of FS)
Range: 2.00 to 19.99 V
Accuracy: ±(1.2% of indication +0.1% of FS)
Range: 2.0 to 1999.9 V
Accuracy: ±(1.2% of indication +0.1% of FS0
Ac voltage
Range: 0.00 to 0.299 V displayed as 0.000 V
0.030 to 1.399 V
(Frequency range: 30 to 500 Hz)
Range: 1.40 to 13.99 V
Accuracy: ±(1.6% of indication +0.3% of FS)
Range: 14.0 to 140.0 V
Accuracy: ±(1.6% of indication +0.3% of FS)
Frequency
Range: 40.00 to 999.9 Hz
(Voltage level: 1.0 to 400 V p-p)
Accuracy: ±0.1% of FS
Resolution: 0.1% Hz
Range: 0.000? to 0.099 kΩ
Resistance
Accuracy: ±(2.0% of indication +1.0% of FS)
Range: 0.100 to 19.99 kΩ
Accuracy: ±(2.0% of indication +0.0% of FS)
Range: 20.00 to 999.9 kΩ
Accuracy: ±(3.0% of indication +0.0% of FS)
Dc current
Range: 0.000 to 2.000 A
Accuracy: ±(1.5% of indication +0.4% of FS)
Ac current
Range: 0.000 to 1.400 A
(Frequency range: 30 to 500 Hz)
Accuracy: ±(1.8% of indication +0.8% of FS)
1Maximum
error 7% up to 1000 A, increasing linearly to 8% at 1500 A.
EQUIPMENT REQUIREMENTS
calibration procedure. This equipment is issued with Secondary Transfer Calibration
Standards Set AN/GSM-287. Alternate items may be used by the calibrating activity. The
items selected must be verified to perform satisfactorily prior to use and must bear evidence
of current calibration. The equipment must meet or exceed the minimum use specifications
standard and TI.
|
|