Page Title: Figure 3. Pressure and vacuum - equipment setup (25 psi).
Back | Up | Next

Click here for thousands of PDF manuals

Google


Web
www.tpub.com

Home


   
Information Categories
.... Administration
Advancement
Aerographer
Automotive
Aviation
Construction
Diving
Draftsman
Engineering
Electronics
Food and Cooking
Logistics
Math
Medical
Music
Nuclear Fundamentals
Photography
Religion
   
   

 

Back
Figure 2. Pressure - equipment setup (1000/10,000 psi).
Up
TB-9-4910-555-35 Calibration Procedures for Simplified Test Equipment for Internal Combustion Engine (STE/ICE-R) Manual
Next
Vacuum

TB 9-4910-555-35
(8) Operate hydraulic pressure standard to apply 1000 psi.
TI will indicate
between 960 and 1040.
(9) Connect equipment as shown in CONNECTION A (fig. 3).
(10) Set TEST SELECTOR switches to 49.
CONNECTION A
CONNECTION B
Figure 3. Pressure and vacuum - equipment setup (25 psi).
(11) Operate pneumatic pressure standard to apply 25 psi and reduce to 0 three times.
(12) With pressure at 0, press and hold TEST pushbutton until CAL appears on
display. Release TEST pushbutton.
(13) Wait for offset value to appear on display. Offset value will be between -4 and +4.
(14) Press and release TEST pushbutton.
(15) Operate pneumatic pressure standard to apply 5 psi. TI will indicate between
4.5 and 5.5.
(16) Operate pneumatic pressure standard to apply 25 psi. TI will indicate between 24
and 26.