TB-9-4910-555-35 Calibration Procedures for Simplified Test Equipment for Internal Combustion Engine (STE/ICE-R) Manual
Next
Figure 2. Pressure - equipment setup (1000/10,000 psi).

TB 9-4910-555-35
12. RPM
a. Performance Check
(1) Connect function/arbitrary generator to TI J3 using cable W3 and W4.
(2) Set TEST SELECT switches to 60.
(3) Press and release TEST pushbutton. Display will indicate UEH.
(4) Set TEST SELECT switches to 01.
(5) Press and release TEST pushbutton. Display will indicate 1.
(6) Set TEST SELECT switches to 10.
(7) Press and release TEST pushbutton. Display will indicate 0.
(8) Adjust function/arbitrary generator for square wave (positive pulse) output of 7
to 13 V and 5000 RPM indication on TI display. Function/arbitrary generator will indicate
between 165.84 and 167.50 Hz.
(9) Repeat technique of (4) through (8) above for indications listed in table 5.
Function/arbitrary generator will indicate within limits specified.
Test instrument
Function/arbitrary generator indications
indications
(Hz)
(RPM)
Min
Max
4500
149.16
150.83
4000
132.50
134.16
3500
115.83
117.49
3000
99.17
100.83
2500
82.50
84.16
2000
65.83
67.49
1500
49.16
50.83
1000
32.50
34.16
500
15.83
17.49
300
9.16
10.83
100
2.50
4.16
50
0.83
2.49
b. Adjustments. No adjustments can be made.