![]() ![]() TB 9-4910-555-35
IDENTIFICATION AND DESCRIPTION
This bulletin provides instructions for the
calibration of Simplified Test Equipment for Internal Combustion Engine (STE/ICE-R).
TM 9-4910-571-12&P and TM 9-4910-571-34&P were used as the prime data sources in
compiling these instructions. The equipment being calibrated will be referred to as the TI
(test instrument) throughout this bulletin.
a. Model Variations. The STE/ICE-R is reprogrammable.
b. Time and Technique. The time required for this calibration is approximately 4
hours, using the dc/low physical techniques.
a. Forms, records, and reports required for calibration personnel at all levels are
prescribed by TB 750-25.
b. Adjustments to be reported are designated (R) at the end of the sentence in which
they appear. When adjustments are in tables, the (R) follows the designated adjustment.
Report only those adjustments made and designated with (R).
pertain to this calibration are listed in table 1.
Test instrument
parameters
Performance specifications
Dc voltage1
Range: 0 to .49 V dc
Accuracy: +6% of FS
Range: .50 to 4.49 V dc
Accuracy: +3% of FS
Range: 4.50 to 45 V dc
Accuracy: +1% of FS
Dc current1
Range: 0 to 1500 A
Accuracy: (+7.0 to 8.0% of input) + (±2% of FS)2
Range: 0 Ω to 4500 Ω
Resistance1
Accuracy: (+3.0% of indication) + (±0.2% of FS)
Range: 0 Ω to 40,000 Ω
Accuracy: (+3% of indication) + ( ±0.5% of FS)
Frequency
Range: 40 to 500 Hz
Accuracy: (±7% of input)+ (±0.65% of FS)
RPM
Range: 50 to 5000 RPM
Accuracy: ±0.5% of FS
See notes at end of table.
|
|