Click here to make tpub.com your Home Page

Page Title: SECTION 1 IDENTIFICATION AND DESCRIPTION
Back | Up | Next

Click here for thousands of PDF manuals

Google


Web
www.tpub.com

Home


   
Information Categories
.... Administration
Advancement
Aerographer
Automotive
Aviation
Construction
Diving
Draftsman
Engineering
Electronics
Food and Cooking
Logistics
Math
Medical
Music
Nuclear Fundamentals
Photography
Religion
   
   

 

Back
CALIBRATION PROCEDURE FOR DEPTH GAGES (GENERAL)
Up
TB-9-4220-215-35 Calibration Procedure for Depth Gages (General) Manual
Next
SECTION III CALIBRATION PROCESS FOR DEPTH GAGES 0-460 FSW AND PRESSURE GAGES -5 TO 30 PSI

TB 9-4220-215-35
SECTION 1
IDENTIFICATION AND DESCRIPTION
1. Test Instrument Identification.
This bulletin provides instructions for the
calibration of Depth Gages (General). Calibration and Repair Requirements Worksheets,
DA Form 3758, and the manufacturers' manuals were used as the prime data sources in
compiling these instructions. The equipment being calibrated will be referred to as the TI
(test instrument) throughout this bulletin.
a. Model Variations. Model variations are described in appendix.
b. Time and Technique. The time required for this calibration is approximately 2
hours, using the physical technique.
2. Forms, Records, and Reports. Forms, records, and reports required for calibration
personnel at all levels are prescribed by TB 750-25.
3. Calibration Description.  TI parameters and performance specifications which
pertain to this calibration are listed in appendix.
SECTION II
EQUIPMENT REQUIREMENTS
4. Equipment Required. Table 1 identifies the specific equipment to be used in this
calibration procedure.  This equipment is issued with Secondary Transfer Calibration
Standards Set AN/USM-286. Alternate items may be used by the calibrating activity. The
items selected must be verified to perform satisfactorily prior to use and must bear evidence
of current calibration. The equipment must meet or exceed the minimum use specifications
listed in table 1. The accuracies listed in table 1 provide a four-to-one ratio between the
standard and the TI.
5. Accessories Required.  The accessories required for this calibration are common
usage accessories, issued as indicated in paragraph 4 above, and are not listed in this
calibration procedure.
Table 1. Minimum Specifications of Equipment Required
Manufacturer and model
Common name
Minimum use specifications
(part number)
PNEUMATIC PRESSURE
Range: 0 to 210 psi
(MIS-30859)
STANDARD
Accuracy: ±0.06 psi
2

Privacy Statement - Press Release - Copyright Information. - Contact Us