Table 5. Ratio Accuracy
Up
TB-9-1430-201-50 Voltage Divider Beckman Models T-10-A Series and TA-1001; ESI Models DP211 DP311 DV411 and DV412; and General Radio Model 1454A Manual
Next
Figure 2. Ratio accuracy (ESI and GR models) - equipment setup

TB 9-1430-201-50
a. Personnel should become familiar with sections I, II, and IV before beginning the
calibration.
b. Items of equipment used in this procedure are referenced within the text by common
name and item identification number as listed in tables 2 and 3. For the identification of
equipment referenced by item numbers prefixed with A, see table 2, and for prefix B, see
Figure 1. Ratio accuracy (Beckman Models) - equipment setup.
11. Input Resistance
a. Performance Check
(1) Connect resistance bridge (A2) UNKNOWN terminals between TI IN and COM
terminals, using Kelvin clips (B4).
(2) Measure TI input resistance. Resistance will indicate within ±0.05 percent of
input resistance specified for TI.
a. Adjustments. No adjustments can be made.
7