![]() ![]() TB 9-1430-201-50
Table 2. Minimum Specifications of Equipment Required - Continued
Minimum use
Manufacturer, model
Item
Common name
specifications
and part number
A4
VOLTAGE STANDARD Range: 12 V dc
John Fluke, Model 332A or
332B/AF (7911393)
A5
LEAD COMPENSATOR Resolution: 0.1 milliohm
ESI, Model 875B (7910539-2)
Table 3. Accessories Required
Item
Common name
Description and part number
B1
LEAD
24-in., No. 18 with single banana plug terminations
1
(red) (7907497)
B2
LEAD1
24-in., No. 18 with single banana plug terminations
(black) (7907498)
B3
SHIELDED CABLE2
Shielded pair No. 18 AWG solid conductor (MIS-10312)
B4
KELVIN CLIPS
Suitable for use with resistance bridge (A2) (7910524)
1Three
required.
2Six
lengths, approximately 24 inches each, required.
SECTION III
CALIBRATION PROCESS FOR
BECKMAN, MODELS T-10-A SERIES AND TA-1001
NOTE
Unless otherwise specified, verify the results of each test and
take corrective action whenever the test requirement is not
met before continuing with the calibration.
6. Preliminary Instructions
calibration.
b. Items of equipment used in this procedure are referenced within the text by common
name and item identification number as listed in tables 2 and 3. For the identification of
equipment referenced by item numbers prefixed with A, see table 2, and for prefix B, see
7. Input Resistance
a. Performance Check
(1) Connect resistance bridge (A2) UNKNOWN terminals to TI input terminals,
using Kelvin clips (B4).
(2) Resistance bridge will indicate within limits specified for TI models listed in
4
|
|
Privacy Statement - Press Release - Copyright Information. - Contact Us |