Figure 2. DC Power Supplies Adjustment Locations.
Up
TB-55-4920-429-35 Calibration Procedure for Advanced Flight Control System Line Set P/N 145G0009-1 Manual
Next
Table 7.0 Analog Meter

TB 55-4290-429-35
10.
Analog Meter.
a.
performance Check.
NOTE
TI analog meter indicates -2 to +2 volts or -20 to
+20 volts.
(1) Observe that digital multimeter (A3) leads (B4) and
(B5) are still connected to the TI HI/LO binding posts.
(2)
Observe that TI METER SOURCE switch is still in
the SELF TEST position.
Set TI 115 vac and 28 vdc POWER switches to ON.
(3)
Digital multi-
(4)
Set TI SELF TEST switch to -7 vdc.
meter will indicate between -6.0 and -8.0 vdc.
Observe indication
on TI analog meter for -7.0 vdc ±1.0 vdc.
(5) Set TI SELF TEST switch to each switch position as
listed in table 7.
Digital multimeter and TI analog meter will
indicate within the limits specified.
(6) Set TI SELF TEST and METER SOURCE switches to OFF.
(7) Disconnect digital multimeter leads from TI binding
posts.
Adjustments.
No adjustments can be made.
b.
18

Privacy Statement - Press Release - Copyright Information. - Contact Us