Figure 2. DVM and Bite Voltages Adjustment Location.
Up
TB-55-4920-428-35 Calibration Procedure for CH-47 Integrated Lower Control Actuator (ILCA) Bench Test Set P/N 145GS278-1 Manual
Next
Position TI controls

TB 55-4920-428-35
(21) Set TI BITE switch to -12 v position. Digital multimeter will indicate between -11.5 and
-12.5 vdc. If not, perform b(4) below as shown in figure 2.
(22) TI DVM will indicate between -11.50 and -12.50.
(23) Set TI POWER to OFF.
(24) Disconnect digital multimeter lead from TI terminal board TB 4-12 (HI) and reconnect to TB
4-11 (HI).
(25) Set TI BITE switch to + 12 v position.
(26) Set TI POWER to ON.
(27)  Digital multimeter will indicate between 11.5 and 12.5 vdc. If not, perform b(5) below as
shown in figure 1.
(28) TI DVM will indicate between 11.50 and 12.50.
(29) Set TI POWER to OFF.
(30) Disconnect digital multimeter lead from TI terminal board TB 4-11 (HI) and reconnect to TB
4-6 (HI).
(31) Set TI BITE switch to 5 v position.
(32) Set TI POWER to ON.
(33) Digital multimeter will indicate between 4.7 and 5.3 vdc. Observe indication on TI DVM
between 4.70 and 5.30.
(34) Set TI POWER to OFF.
(35) Disconnect digital multimeter leads (B6) from TI terminal board TB 4-6(HI) and TB 4-
5 (LOW) and reconnect to EXTERNAL METER jacks, using leads (B7) and (B8).
(36) Connect Electronic Counter (A3) leads (B3) to the digital multimeter leads.
(37) Set digital multimeter RANGE switches to 20 vac.
(38) Position TI controls (a) thru (c) as follows:
(a) EXTERNAL METER SELECT to REF and SYSTEM SELECT to BITE.
(b) BITE to 13 RMS.
(C) POWER to ON.
(39) Digital multimeter will indicate between 12.5 and 13.5 vac. Electronic Counter will indicate
between 395 and 405 Hz. If not, perform b(6) below as shown in figure 2.
(40) TI DVM will indicate between 12.50 and 13.50.