![]() ![]() TB 55-4920-389-34
DEPARTMENT OF THE ARMY TECHNICAL BULLETIN
CALIBRATION PROCEDURE FOR TEST STAND, IGNITION MAGNETO HYDRAULIC
TECHNOLOGY, INC. MODELS 6990A
Headquarters, Department of the Army, Washington, D. C.
13 May 1977
Section I. IDENTIFICATION AND DESCRIPTION
1-1. Test Instrument Identification.
This bulletin provides instructions for the calibration of Ignition Magneto Test Stand. The Ignition Magneto T.S. will
be referred to as the "test instrument" throughout this bulletin.
Nomenclature: Ignition Magneto Test Stand
FSN:
4920-00-
Manufacturer: Hydraulic Technology Inc.
Model:
6990 and 6990A
Model variation:
Only difference is the size.
Technique:
DC and low frequency
1-2. Calibration Description.
Test instrument parameters and performance specifications which pertain to this calibration are listed in table 1-1.
Table 1-1. Calibration Description
Test instrument parameters
Performance specifications
RPM Indicator (dual range)
Range: 0-500, 0-5000 RPM
Accuracy: +3.0% of F.S. for each range
AC Ammeter (dual range)
Range: 0-2.5, 0-5.0 amps a.c.
Accuracy: +3.0% of F.S. for each range