Section III. PRELIMINARY OPERATIONS
Up
TB-55-4920-343-35-1 Calibration Procedure for Simulator Self-Tuning Vibration Absorber System P/N 114G1409-1 and 114G5090-1 Manual
Next
Ripple Voltage.

TB 55-4920-343-35/1
e. Set S1 switch to ON. Phase indicator must light.
f. Allow time for equipment to warm up and stabilize.
Section IV. CALIBRATION PROCESS
NOTE
Unless otherwise specified, verify the results of each test and take corrective action whenever
the test requirements are not met before continuing with the calibration.
4-1.
Operating Voltage.
a. Performance Check.
(1) Connect equipment as shown in figure 4-1.
(2) Set multimeter (Al) range switch setting to the 50 vdc position.
Figure 4-1. Operating voltage-equipment setup.
5