Page Title: Table 4. RESISTANCE Gage Performance Limits
Back | Up

Click here for thousands of PDF manuals

Google


Web
www.tpub.com

Home


   
Information Categories
.... Administration
Advancement
Aerographer
Automotive
Aviation
Construction
Diving
Draftsman
Engineering
Electronics
Food and Cooking
Logistics
Math
Medical
Music
Nuclear Fundamentals
Photography
Religion
   
   

 

Back
Figure 3. AIR FLOW gage performance check, equipment setup
Up
TB-3-6665-315-50 Calibration Procedure for Indicator Resistance Inhilation and Exhaltion Q213 Manual

Table 4. RESISTANCE Gage Performance Limits
Scale Major
Acceptable
Division Indication
Manometer Indications
(Centimeters of Water)
(Inches of water)
Max.
Min.
0.866
0.709
2
1.653
1.496
4
2.441
2.283
6
3.228
3.071
8
4.016
3.858
10
the TI does not indicate within the limits specified in
10. Final Procedure. a. Deenergize and disconnect
paragraph 8a or in table 4, annotate the affix DA Form
all equipment.
2417 (Unserviceable or Limited Use) tag.
b.  In accordance with TM 38-750, annotate and
affix DA Label 80 (US Army Calibration System). When
By Order of the Secretary of the Army:
BERNARD W. ROGERS
General, United States Army
Official:
Chief of Staff
PAUL T. SMITH
Major General, United States Army
The Adjutant General
DISTRIBUTION:
To be distributed in accordance with DA Form 12-34A requirements for Calibration Procedures Publications.
U.S. GOVERNMENT PRINTING OFFICE: 1977-765010/530
6