![]() ![]() TB 11-6625-2937-35
b. Adjustment.
reinstall covers on control-interface unit.
b. In accordance with TM 38-750, annotate
(1) Adjust power supply output voltage
and affix DA Label 80 (US Army Calibrated Instrument).
adjustment (ADJ POT) for a voltage indication
When the equipment receives limited
of +28.0+0.2 V on the multimeter.
or special calibration, annotate and affix DA Label
(2) Set POWER ON/OFF switch to OFF.
163 (US Army Limited or Special Calibration).
(3) Remove multimeter from terminals 3
When the equipment cannot be adjusted within
(+) and 4 (-) of power supply AlPS3.
tolerance, annotate and affix DA Form 2417 (US
Army Calibration System Rejected Instrument).
13. Final Procedure
a. Deenergize and disconnect all equipment and