TB-11-6625-2932-35 Calibration Procedure for Cathode for Cathode Relay Tester Siemens & Halske 9 Tmse 109 Manual
Next
Section II. EQUIPMENT REQUIREMENTS

TB 11-6625-2932-35
DEPARTMENT OF THE ARMY TECHNICAL BULLETIN
CALIBRATION PROCEDURE
FOR
CATHODE RAY RELAY TESTER
SIEMENS & HALSKE 9Tmse 109
Headquarters, Department of the Army, Washington DC
4 August 1980
REPORTING OF ERRORS
You can help improve this publication by calling attention to errors and by
recommending improvements and stating your reasons for the recom-mendations. Your
letter or DA Form 2028 (Recommended Changes to Publication or Blank Forms) should
be mailed direct to Commander, US Army Communications and Electronics Materiel
Readiness Command, ATTN: DRSEL-ME-MQ, Fort Monmouth, NJ 07703. A reply will be
furnished direct to you.
Paragraph
Page
SECTION I. IDENTIFICATION AND DESCRIPTION
Test instrument identification ....................................................................................
Calibration data card ................................................................................................
Calibration description ..............................................................................................
II. EQUIPMENT REQUIREMENTS
Equipment required ..................................................................................................
Accessories required ................................................................................................
III. PRELIMINARY OPERATIONS
Preliminary instructions ............................................................................................
Equipment setup.......................................................................................................
SECTION IV. CALIBRATION
Deflection voltage .....................................................................................................
Frequency divider circuit...........................................................................................
Power voltages and current ......................................................................................
Final procedure.........................................................................................................
Section I. IDENTIFICATION AND DESCRIPTION
b. Time and Technique. The time required to perform
1.
Test Instrument Identification. This bulletin
this calibration is approximately 2 hours using the dc low
provides instructions for the calibration of Cathode Ray
frequency technique.
Relay Tester, Siemens Halske 9Tmse109. The cathode
ray relay tester being calibrated will be referred to as the
2.  Calibration Data Card. a. Forms, records, and
TI (test instrument) throughout the bulletin.
a. Model Variations. This procedure pertains to only
reports required for calibration personnel at all levels are
one model. Some instruments have trimmer capacitors
prescribed by TM 38-750. DA Form 2416
C5c and C5d and some do not, according to the balance
of the circuitry of resistors R1 and R22 and capacitor
assembly C5.
1