![]() ![]() volts and back to 115 volts.
Frequency counter will
(8) Adjust FREQUENCY GHZ control to
indicate within 0.008 percent of indication recorded in
positions listed in table 4.
If frequency counter
(6) above.
indications are not within limits specified, perform b
below.
Table 4. Frequency Dial Check
Frequency counter indications (GHz)
Test instrument
FREQUENCY GHz
control settings
Min
Max
4.50
9.955
9.045
5.50
10.945
11.055
6.50
12.935
13.065
7.50
14.925
15.075
8.40
16.716
16.884
b
Adjustments (fig. 1)
Figure 1. Signal source--top view.
4
|
|