Table 5. Time-Division Accuracy
Up
TB-11-6625-2734-50 Calibration Procedure for TDR/ Sampler Plug-In TD-1160(P)/U Manual

TB 11-6625-2734-50
(4) Set TI switch S901 (fig. 2) from off (rear
(7) Depress TI HIGH RESOLUTION switch to
position) to normal (center position). If trace shifts on
out position and turn SCAN control fully clockwise.
oscilloscope crt, perform b below.
(8) Set TI TIME-DISTANCE multiplier switch to
(5) Set TI switch S901 to normal (center)
X.1 and TIME/DIV switch to .2 ns.
position.
(9) Position  leading  edge  of  trace  on
b. Adjustments. Adjust R945 (fig. 2) for no trace
oscilloscope crt vertical graticule centerline, using TIME-
shift as S901 is switched from off to normal position.
DISTANCE control.
17. Risetime. a. Performance Check
(10) Depress TI LOCATE switch to out position.
(1) Connect pulse generator (A4) to sampling
If center of trace on leading edge is not bright, perform
b(4) below.
head (A7), using cable (B6).
b. Adjustments.
(2) Set TI TIME/DIV switch to 20 ps and TIME-
DISTANCE control to display pulse on oscilloscope (A3)
(1) Adjust R594 (fig. 1) to position pulse on
crt.  Adjust mV/DIV VARIABLE control for suitable
oscilloscope crt vertical graticule centerline.
display.
(2) Adjust R592 (fig. 1) to position pulse on
(3) Measure rise time, using standard rise-time
oscilloscope crt vertical graticule centerline.
technique. Rise time will be 35 picoseconds or less.
(3) Adjust R618 (fig. 2) for one sweep in 50 or
b. Adjustments. No adjustments can be made.
more seconds.
18. Final Procedure. a. Deenergize and disconnect all
(4) Adjust R677 (fig. 2) to center bright portion
equipment.
of trace on leading edge of trace.
b. In accordance with TM 38-750, annotate and
16. Correction Memory. a. Performance Check
affix DA Label 80 (US Army Calibration System). When
(1) Depress TI LOCATE switch to in position.
the TI cannot be adjusted within tolerance, annotate and
(2) Connect termination (B10) and cable (B7) to
affix DA Form 2417 (Unserviceable or Limited Use) tag.
sampling head (A7)
(3) Position  trace  on  horizontal  graticule
centerline on oscilloscope (A3) crt.
9