SECTION III PRELIMINARY PROCEDURE
Up
TB-11-6625-2715-50 Calibration Procedure for Ammeter Simpson 2100 Manual

TB 11-6625-2715-50
SECTION IV
CALIBRATION
1-9. Current Calibration. a. Performance Check.
setup and adjust current shunt for 30 amp reading.
(8) Adjust current supply for a TI full scale
(1) Adjust Current Supply A3 and Decade
indication.
Resistor A4 for a TI full scale indication.
(9) Observe  that  the  differential  voltmeter
(2) Observe that the DC Differential Voltmeter
indicates between 294 and 30.6 millivolts.
indicates between 14.7 and 15.3 millivolts.
b. Adjustment. No adjustments can be made.
(3) Remove positive lead from 1.5 ampere
connector and connect to TI 3.0 ampere connector.
1-10. Final Procedure. a. Deenergize and disconnect
(4) Adjust current supply and decade resistor
all equipment.
for a TI full scale indication.
b. In accordance with TM 38-750, annotate and
(5) Observe  that  the  differential  voltmeter
affix DA Label 80 (US Army Calibration System). When
indicates between 24 and 30.6 millivolts.
the TI cannot be adjusted within tolerance, annotate and
(6) Remove positive lead from 3.0 ampere
affix Form 2417 (Unserviceable or Limited Use) (Red
connector and connect to TI 30 ampere connector.
Tag).
(7)  Remove decade resistor from equipment
5