Page Title: Sweep Length
Back | Up | Next

Click here for thousands of PDF manuals

Google


Web
www.tpub.com

Home


   
Information Categories
.... Administration
Advancement
Aerographer
Automotive
Aviation
Construction
Diving
Draftsman
Engineering
Electronics
Food and Cooking
Logistics
Math
Medical
Music
Nuclear Fundamentals
Photography
Religion
   
   

 

Back
Horizontal Bandwidth
Up
TB-11-6625-2713-50 Calibration Procedure for Oscilloscope AN/USM-296A Manual
Next
Table 6. Normal and Delayed Sweep Rates

TB 11-6625-2713-50
b. Adjustments.
Adjust C25 (fig. 7) until crt indicates at least 4.25 divisions of
deflection (R).
Figure 7. Horizontal sweep, adjustments and test point locations.
18. Sweep Length
a. Performance Check
(1) Set TI controls as listed in (a) through (c) below:
(a) AUTO-TRIG SINGLE switch to AUTO.
(b) Horizontal mode switch to NORM.
(c) SOURCE switch to INT.
(2) Position trace to start of left vertical graticule line on crt. If sweep length is not
10.5 divisions, perform b below (R).
b. Adjustments. Adjust X1 GAIN CAL A1A2R22 (fig. 7) for sweep length of 10.5
divisions.
19. Sweep Calibration
a. Performance Check
(1) Connect time-mark generator (A6) to TI CH1 INPUT, using cable (b4).
15