Click here to make tpub.com your Home Page

Page Title: Horizontal Bandwidth
Back | Up | Next

Click here for thousands of PDF manuals

Google


Web
www.tpub.com

Home


   
Information Categories
.... Administration
Advancement
Aerographer
Automotive
Aviation
Construction
Diving
Draftsman
Engineering
Electronics
Food and Cooking
Logistics
Math
Medical
Music
Nuclear Fundamentals
Photography
Religion
   
   

 

Back
Horizontal Sweep
Up
TB-11-6625-2713-50 Calibration Procedure for Oscilloscope AN/USM-296A Manual
Next
Sweep Length

TB 11-6625-2713-50
(2) Adjust TRIG LEVEL CENTERING ADJ A3A5R6 (fig. 5) for +0.65 volt
indication on digital voltmeter (R).
(3) Adjust A3A5R82 (fig. 5) for -0.25 volt indication on digital voltmeter (R).
(4) Adjust CRT BIAS ADJ A1A2R12 (fig. 6) for 50 volts indication on digital
voltmeter (R).
(5) Adjust A1A2R24 (fig. 6) for no shift of spot as X10 VERNIER control is pulled
out and pressed in (R).
Figure 6. Horizontal amplifier, test point location.
17. Horizontal Bandwidth
a. Performance Check
(1) Connect oscillator (A4) to TI EXT HOR input, using cable (B4).
(2) Set both TI SOURCE switches to EXT.
(3) Adjust oscillator frequency for 50 kHz and amplitude for 6 divisions of deflection
on crt.
(4) Adjust oscillator frequency until crt displays 4.25 divisions of deflection.
If
oscillator does not indicate 5 MHz or greater, perform b below.
14

Privacy Statement - Press Release - Copyright Information. - Contact Us