![]() ![]() TB 11-6625-2713-50
(2) Rotate CH1 POSITION control fully counterclockwise and then fully clockwise.
If negative and positive portion of display is not equal distance above and below horizontal
graticule centerline, perform b below.
below horizontal graticule centerline.
Figure 3. Vertical amplifier adjustments (subassembly 7963).
12. Attenuator and Input Capacitance
a. Performance Check
(1) Connect square-wave generator (A5) OUTPUT to TI CH1 INPUT, using cable
and termination supplied with square-wave generator, standardizer, and adapters (B6, B1,
and B2).
(2) Set TI CHI VOLTS/DIV switch to .005.
(3) Adjust square-wave generator frequency for 1 kHz (kilohertz) and amplitude for
4 divisions of vertical display on crt.
(4) Adjust time base for convenient display on crt.
(5) Adjust standardizer for optimum squarewave display.
(6) Repeat technique of (2) and (3) above for TI VOLTS/DIV switch settings listed in
appropriate adjustments listed.
10
|
|