DEPARTMENT OF THE ARMY TECHNICAL BULLETIN
Up
TB-11-6625-2713-50 Calibration Procedure for Oscilloscope AN/USM-296A Manual
Next
Test instrument VOLTS/DIV switch settings

TB 11-6625-2713-50
CHANGE 1
DEPARTMENT OF THE ARMY TECHNICAL BULLETIN
CALIBRATION PROCEDURE FOR
OSCILLOSCOPE AN/USM-296A
(NSN 6625-00-115-9201)
Headquarters, Department of the Army, Washington, DC
17 June 1980
TB 11-6625-2713-50, 29 December 1978, is changed as follows:
Page 4, paragraph 7a(6), line 1. "CH 1" is changed to read: CH 2.
Paragraph 8a(6), line 2. "CH 1" is changed to read: CH 2.
Paragraph 8a(8), line 1. "CH 1" is changed to read: CH 2.
Paragraph 9a(2), line 2. "CH 1" is changed to read: CH 2
Paragraph 9a(3), line 1. "CH 1" is changed to read: CH 2.
Page 5, paragraph 9a(5) and (6) are superseded as follows:
(5) Set TI VERTICAL MODE to CH 2.
(6) Repeat steps (*3) and (4) above for channel 2.
(7) Connect digital voltmeter (A3) to junction of A2A3R9 and A2A3R20 (fig. 2) and
chassis ground. If digital voltmeter does not indicate zero voltage ±50 mV, perform b(1)
below.
(8) Set TI VERTICAL MODE to CH 1. If digital voltmeter does not indicate zero
voltage ±50 mV, perform b(2) below.
(9) Set TI VERTICAL MODE to CH 1.
PIN: 015255-001