Helix Feedback Amplifier Gain
Up
TB-11-6625-2402-35-1 Calibration Procedure for Sweep Generators AN/USM-308(V) AN/USM-308(V)1 AN/USM-308(V)2 and AN/USM-308(V)3 Manual
Next
Marker (High End)

(c) MANUAL SWEEP control fully clock-
(f) MARKER AMP adjustment fully
wise.
clockwise.
(d) FUNCTION START STOP pushbut-
(2) Set calibrator FUNCTION SELECTOR
ton pressed,
switch to SWP REF AM.
(e) All AMPLITUDE MOD pushbuttons
(3) Adjust UUT MARKER 1 START/CW
released.
control for maximum dc deflection on oscillo-
(2) Set calibrator FUNCTION SELECTOR
scope (l-4), UUT MI pointer indicates 3 V on
switch to SWP REF AM.
calibrator scale.
(3) Short UUT EXT FM connector with
(4) Position UUT controls as follows:
adapter (l-9) and test lead (1-10).
(a) AMPLITUDE MOD MARK 1 push-
button released.
(4) Multimeter indicates between 72.9 and
(b) AMPLITUDE MOD MARK 2 push-
73.1 vdc.
button pressed.
(5) Press UUT FUNCTION MARKER
(5) Adjust UUT MARKER 2 STOP control
SWEEP pushbutton.
for maximum dc deflection on oscilloscope. UUT
(6) Multimeter indicates between 72.9 and
M2 pointer indicates 3 V on calibrator scale.
73.1 Vdc.
b. Adjustments.
(7) Set UUT SWEEP SELECTOR switch
to Cw.
(1) Repeat performance check steps (1) and
(2) .
(8) Multimeter indicates between 72.9 and
(2) Adjust UUT R18 (fig. 3-5) for maxi-
73.1 vdc.
mum dc deflection on oscilloscope.
(9) Press UUT FUNCTION AF pushbutton.
(3) Repeat performance check step (4).
(10) Multimeter indicates between 72.9 and
(4) Adjust UUT R2 (fig. 3-5) for maximum
73.1 vdc.
dc deflection on oscilloscope.
(11 ) Press UUT FUNCTION EXT FM push-
(5) Repeat performance check.
button.
(12) Allow 3 minutes for circuit to stabilize.
3-9. Frequency Control (High End)
Multimeter indicates between 72.9 and 73.1 vdc.
a. Performance Check.
(13) Remove adapter (1-9) and cable (l-
(1) Position UUT controls as
follows:
10) from EX.T FM connector.
(a) START/CW, MARKER
1 START/
CW, MARKER 2 STOP and STOP/
F controls
b. Adjustments.
to 73 V on calibrator scale.
(1) Repeat performance check steps (1)
( b ) SWEEP SELECTOR
switch to
through (3).
MANUAL.
(2) Adjust UUT STOP HF R24 (fig. 3-6)
for multimeter indication of 73.0 vdc.
(3) Press UUT FUNCTION MARKER
SWEEP pushbutton.
(4) Adjust UUT MARKER STOP HF R30
(fig. 3-6) for multimeter indication of 73.0 vdc.
(5) Set UUT SWEEP SELECTOR switch
to Cw.
(6) Adjust UUT MARK START HF R22
(fig. .3-6) for multimeter indication of 73.0 vdc,
(7) Press UUT FUNCTION
F pushbutton.
ton.
(8) Adjust UUT
F CW R19 (fig. 3-6)
for multimeter indication of 73.0 vdc.
(9) Press UUT FUNCTION EXT FM push-
button Allow 3 minutes for circuit to stabilize.
18