Click here to make tpub.com your Home Page

Page Title: Table 2-2. Resistance Check (DCVM Connector)
Back | Up | Next

Click here for thousands of PDF manuals

Google


Web
www.tpub.com

Home


   
Information Categories
.... Administration
Advancement
Aerographer
Automotive
Aviation
Construction
Diving
Draftsman
Engineering
Electronics
Food and Cooking
Logistics
Math
Medical
Music
Nuclear Fundamentals
Photography
Religion
   
   

 

Back
SECTION II C A L I B R A T I O N  OF  CALIBRATORS
Up
TB-11-6625-2402-35-1 Calibration Procedure for Sweep Generators AN/USM-308(V) AN/USM-308(V)1 AN/USM-308(V)2 and AN/USM-308(V)3 Manual
Next
Figure 2-1. Typical oscilloscope waveform

(4) Set UUT FUNCTION SELECTOR
switch to settings listed in table 2-3. Multimeter
indicates with specified limits.
(5) Set UUT FUNCTION SELECTOR
switch to HELIX (10:1) /HELIX (1:1AC).
(6) Reverse banana plugs on multimeter.
(7) Multimeter indicates between 2K ohms
and 4K ohms.
(8) Set UUT FUNCTION SELECTOR
switch to - 6.3V
(9) Connect multimeter to UUT test points
listed in table 2-4, with 2 test leads (1-9) and
2 probes (1-10). Multimeter indicates within
specified limits.
NOTE
Install UUT in sweep oscillator (1-5).
b. Adjustments. No adjustments can be made.
2-3. Voltage and Waveforms
Table 2-2. Resistance Check (DCVM Connector)
a. Performance Check.
(1) Connect voltmeter (1-2) to UUT DC-
VM jack with cable assembly (1-7).
(2) Connect UUT SCOPE ACVM jack to
oscilloscope channel A input using a connector
(1-8) and cable assembly ( 1-7). Connect cable
assembly (1-7) from connector (1-8) to volt-
meter (1-4).
(3) Turn sweep oscillator (1-5) LINE
switch to STANDBY and allow approximately
1 minute for equipment warmup.
(4) Depress UUT TIME DELAY OVER-
RIDE pushbutton for about 5 seconds.
(5) Turn sweep oscillator LINE switch to
RF.
(6) Set UUT FUNCTION SELECTOR
switch to positions listed in table 2-5 and observe
that the indications are within the prescribed
limits.
b. Adjustments. No adjustments can be made.
Voltmeter  indication
(vol-- .
ts)
UUT FUNCTION SELECTOh
switch position
minimum
maximum
I
SWP REF AM
+40
+36
-300 -300. . . . . . . . . . . . . . . . . . . . . . . . . .
-280
-320
-6.3V-6.3V . . . . . . . . . . .. . . . . . . . . . . . . . . . . . . . . . .
-6.5
-6.1
+ 276V + 275V
+ 294
.
+ 256
+20V/+20V . . . . . . . . . . . . . . . . . . . . . . .
+21
+19
ANODE/START RMP . . . . . . . . . . .
+ 104
+ 116
HELIX FDBK STOP RMP
+ 33.3
+ 37.31
.
HELIX (10:1) /HELIX (1:1AC)
+124 3
+112
COLL(10:1) BLANKING . . . . . .
..
+ 136.5
+118
1
Note this Indicatlon.
2
Add 12.6 volts to the HELIX (10:1) HELIX (1:1 AC) voltage.
3
See figure 2-1 for typical oscilloscope presentation.
12

Privacy Statement - Press Release - Copyright Information. - Contact Us