![]() ![]() TB 11-6625-2402-35/1
(6) Set MANUAL SWEEP control fully
TB 11-6625-2402-35/1, 8 May 1973, is
CW. Record electronic counter indication.
changed as follows:
(7) Subtract value recorded in (5) above
from value recorded in (6) above. Sweep width
Page 12, Table 2-3. In column 2, line 7, add
" * " after "zero".
will be between 0.97 and 1.03 MHz.
(8) Set SCALE switch to X1.
(9) Adjust START/CW control for a
Page 35. Delete tabIes 6-3 and 6-4.
50.00-MHz indication on electronic counter.
Paragraph 6-4 a (5) through (10) is superseded
(10)
Repeat (5) through (7) above
a s follows
except sweep width will be between 9.7 and
10.3 MHz.
(5) Set MANUAL SWEEP control fully
C C W. Record electronic counter indication.