TB-11-6625-238-35 Calibration Procedure for Telephone Carrier System Test Facilities Kit MK-155/TCC Manual
Next
Table 6. Matching Transformer Set, Resistor Accuracy Requirements

Section III. CALIBRATION PROCESS
Personnel should familiarize themselves with the entire procedure before performing calibration.
7. Preliminary Procedure. Remove measuring set,
and allow equipment to warm up for 15 minutes.
matching transformer set, and unassembled parts
Figure 2 Telephone carrier system test facilities kit
container from carrying cases. Connect equipment as
(measuring set) equipment setup
illustrated in figures 2 and 3. Turn on power switches
Figure 3 Transformer chassis, with impedance matching network, physical equipment assembly