Click here to make tpub.com your Home Page

Page Title: VOLTAGE AC PHASE A Meter Calibration Accuracy
Back | Up | Next

Click here for thousands of PDF manuals

Google


Web
www.tpub.com

Home


   
Information Categories
.... Administration
Advancement
Aerographer
Automotive
Aviation
Construction
Diving
Draftsman
Engineering
Electronics
Food and Cooking
Logistics
Math
Medical
Music
Nuclear Fundamentals
Photography
Religion
   
   

 

Back
Preliminary Procedures
Up
TB-11-5995-207-35-2 Calibration Procedure for Montor Electrical Power MX-8570/APS-94D Manual
Next
VOLTAGE DC Meter Calibration Accuracy.

Figure 3. Calibration test setup.
(2) Compare voltage levels indicated on volt-meter and
8. VOLTAGE AC PHASE A Meter Calibration
VOLTAGE AC PHASE C meter of unit under test. The
Accuracy. a. Performance Check.
VOLTAGE AC PHASE C meter should indicate voltage
(1) Energize equipment and allow 5 minutes for
level displayed on voltmeter, + 2 percent.
equipment to warm up and stabilize.
b. Adjustments. No adjustments can be made.
(2) Compare voltage levels indicated on volt-meter and
11.  AMPERE  AC  PHASE  A  Meter  Calibration
VOLTAGE AC PHASE A meter of unit under test. The
Accuracy. a. Performance Check.
VOLTAGE AC PHASE A meter should indicate voltage
level displayed on voltmeter + 2 percent.
(1) Trip 115 VOLT AC CIRCUIT BREAKER of unit under
b. Adjustment. No adjustments can be made.
test.
(2) Arrange ammeter connections for 10 ampere
9.  VOLTAGE  AC  PHASE  B  Meter  Calibration
operation and connect to test jacks J10 and J16 of power
Accuracy. a. Performance Check.
monitor test fixture with test leads.
(1) Connect voltmeter to test jacks J2 and J4 of power
(3) Close 115 VOLT AC CIRCUIT BREAKER of unit
monitor test fixture (fig. 2) with test leads.
under test.
(2) Compare voltage levels indicated on volt-meter and
(4) Compare current levels indicated on ammeter and
VOLTAGE AC PHASE B meter of unit under test. The
AMPERE AC PHASE A meter of unit under test. The
VOLTAGE AC PHASE B meter should indicate voltage
AMPERE AC PHASE A meter should indicate current
level displayed on voltmeter, + 2 percent.
level displayed on ammeter, + 2.1 percent.
b. Adjustments. No adjustments can be made.
12.  AMPERE  AC  PHASE  B  Meter  Calibration
10. VOLTAGE AC PHASE C Meter Calibration
Accuracy. a. Performance Check.
Accuracy. a. Performance Check.
(1) Trip 115 VOLT AC CIRCUIT BREAKER of unit under
(1) Connect voltmeter to test jacks J3 and J4 of power
test.
monitor test fixture with test leads.
5