Figure 2. Preliminary equipment setup.
Up
TB-11-5820-861-35 Calibration Procedure for Automatic Video Corrector Tektronix Model 1440 Manual
Next
Figure 3. Timing and decoder boards - component locations.

SECTION IV.
CALIBRATION PROCESS
minimum chroma (signal level) on 0.5-volt pedestal of
waveform.
NOTE
Unless otherwise specified, verify the
(14) Connect oscilloscope probe to TP4539 (fig.
results of each test and, whenever the test
3). Adjust L4460 and L4456 (fig. 3) for optimum peak
requirement is not met, take corrective
response of the demodulated VIR signal and chroma.
action  before  continuing  with  the
(15) Adjust video signal source (A6) for 10
calibration.
percent APL (average picture level) flat field output. If
APL, as indicated on oscilloscope is not zero, perform
8. Timing Circuits. a. Performance Check.
b(8) below.
(16) Adjust TI PEDESTAL DELAY control for
(1) Connect the oscilloscope (A4) to TI TP3308
minimum aberration on waveform approximately 62
(fig. 3), using probe (B2).
microseconds from leading edge of horizontal sync
(2) If dc level indicated on oscilloscope exceeds
pulse.
+20 millivolts, perform b(l) below.
(17) Adjust TI PEDESTAL START control for
(3) Connect oscilloscope probe to TP3250 (fig.
minimum aberration on waveform approximately 11
microseconds from leading edge of horizontal sync
If width of displayed pulse is not 4 +0.2 microseconds,
pulse.
perform b(2) below.
(18) Adjust video signal source for 0 APL flat
(4) Turn TI PEDESTAL DELAY and START
field output. If APL, as indicated on oscilloscope, is not
controls to midrange.
zero, perform b(8) below.
(5) Connect oscilloscope probe to TP3385 (fig.
3).If width of displayed pulse is not 11 +1 microseconds,
b. Adjustments (1) Adjust L3400 (fig. 3) for oscilloscope
perform b(3) below.
indication of zero volt dc (R).
(6) Connect oscilloscope probe to P3912-3 (fig.
3) (7) Adjust oscilloscope to display horizontal sync pulse
(2) Adjust R3624 (fig.  3) for 4-microsecond
and position leading edge of pulse at second vertical
pulse width (R).
graticule line.
(3) Adjust R3156 (fig.  3) for 11-microsecond
(8) Connect oscilloscope probe to TP3876 (fig.
pulse width (R).
3).If trailing edge of displayed pulse is not 8.5 +0.5
(4) Adjust R3282 (fig. 3) until trailing edge of
microseconds  from  second  vertical  graticule  line,
displayed pulse is 8 microseconds from second vertical
perform b(4) below.
graticule line (R).
(9) Repeat (6) and (7) above.
(5) Adjust R3534 (fig. 3) until leading edge of
(10) Connect oscilloscope probe to TP4141 (fig.
displayed pulse is 17 microseconds from second vertical
3). If leading edge of displayed pulse is not 17 +0.5
graticule line (R).
microseconds from second vertical graticule, perform
(6) Adjust R3438 (fig. 3) until trailing edge of
b(5) below; if trailing edge is not 32 +0.5 microseconds
displayed pulse is 32 microseconds from second vertical
from second vertical graticule, perform b(6) below.
graticule line (R).
(11) Connect oscilloscope probe to TP4139 (fig.
(7) Adjust R3540 (fig. 3) until trailing edge of
3).  f trailing edge of displayed pulse is not 46 +1
displayed pulse is 46 microseconds from second vertical
microseconds  from  second  vertical  graticule  line,
graticule line (R).
perform b(7) below.
(8) Adjust R5404 (fig. 4) for zero APL indication
(12) Connect oscilloscope probe to TP4505 (fig.
on oscilloscope (R).
(13) Adjust oscilloscope to display VIR (vertical
interval reference) signal.  Adjust L4378 (fig.  3) for
6