Order this information in Print

Order this information on CD-ROM

Download in PDF Format

     

Click here to make tpub.com your Home Page

Page Title: Troubleshooting Table 4-7. Analyzer Test Set Circuits
Back | Up | Next

Click here for a printable version

Google


Web
www.tpub.com

Home


   
Information Categories
.... Administration
Advancement
Aerographer
Automotive
Aviation
Combat
Construction
Diving
Draftsman
Engineering
Electronics
Food and Cooking
Math
Medical
Music
Nuclear Fundamentals
Photography
Religion
USMC
   
Products
  Educational CD-ROM's
Printed Manuals
Downloadable Books
   

 

Back
m. Timer  Time  Delay  Circuit.
Up
TM-55-4920-410-13P Vibrex Balancing Kit Part No. D4591 Manual
Next
Analyzer Test Set Circuits (continued)

FO-1 tween 2.5 and 3.5 see, adjust R7 for 3Sec. If specification cannot be met, go to troubleshooting Table 4-7 (Fault No. 6). n. Timer Logic Tests Circuit. (1) Disconnect all test equipment from test set, ex- cept DC power supply as shown in Figure 4-2. (2) Set TRIP TIME -START switch to DOWN and TEST SELECT to AØ UV position. Jumper J1 pins, a, b, g, and set TRIP TIME-START switch to UP. (3)  TRIP-TEST  IN  PROG  indicator  and  TRIP TIME-EARLY TRIP indicator will light. TRIP-TEST IN PROG  indicator  will  go  OFF  and  TRIP  TIME-TEST COMPL indicator will light. If not, see troubleshooting Table 4-8 (Faults No. 7, 8, & 11). (4)  Remove  jumper  from  J1  pin  a  and  set  TRIP TIME START stitch to DOWN. TRIP TIME lamps will TM 55-4920-410-13&P go OFF. If not, see troubleshooting Table 4-8 (Fault No. l0). (5) Set TRIP TIME-START switch to UP. TRIP TIME TEST IN PROG indicator will light. TRIP TIME- TEST COMPL indicator and TRIP TIME-L, ATE TRIP indicator will light and TRIP TIME-TEST IN PROG in- dicator will go off. If not, see troubleshooting Table 4-8 (Faults No. 7, 10, & 11). (6) Set TRIP TIME-START switch to DOWN and set TEST SELECT switch to 3Ø OVB and jumper pin a to  b  and  g.  Set  TRIP  TIME-START  switch  to  UP.  “ (7) TRIP-TEST IN PROG indicator will light and TRIP  TIME-ACCEPT  indicator  will  also  light  when TRIP TIME-START switch is placed to UP. TRIP TIME- TEST IN PROG indicator will go off. TRIP TIME-TEST COMP indicator will light. If not, see troubleshooting Ta- ble 4-8 (Faults No. 7,9, & 11). Set TRIP TIME-START switch to DOWN. Troubleshooting Table 4-7. Analyzer Test Set Circuits 4-11

Privacy Statement - Press Release - Copyright Information. - Contact Us - Support Integrated Publishing