Order this information in Print

Order this information on CD-ROM

Download in PDF Format

     

Click here to make tpub.com your Home Page

Page  Title: SECTION VIII UT PROCESS CONTROL
Back | Up | Next

Click here for a printable version

Google


Web
www.tpub.com

Home


   
Information Categories
.... Administration
Advancement
Aerographer
Automotive
Aviation
Combat
Construction
Diving
Draftsman
Engineering
Electronics
Food and Cooking
Math
Medical
Music
Nuclear Fundamentals
Photography
Religion
USMC
   
Products
  Educational CD-ROM's
Printed Manuals
Downloadable Books
   

 

Back
Figure 1-9b. Checking In-Use Field Indicators
Up
TM-1-1500-335-23 Nondestructive Inspection Methods Manual
Next
Table 1-5.  Reference Standard Metal Travel Tolerances.

T.O. 33B-1-1 SECTION VIII UT PROCESS CONTROL 1.8 ULTRASONIC PROCESS CONTROL REQUIREMENTS. 1.8.1 The Ultrasonic Process Control Requirements. In the ultrasonic inspection process like the eddy current process you must know your equipment and probe or  transducer  or  search  unit  is  functioning  properly.    The  following  information  on  process  control  in ultrasonic  testing  is  designed  to  help  ensure  repeatability  and  accuracy.      All  new  transducers  should  be checked  against  a  reference  transducer  of  the  same  dimension  and  frequency.    Only  transducers  in  use require documentation on serviceability.  For example, you may have 5 of the same transducers in stock but only be using one.  The one in use should be labeled to show the required gain to achieve a 100% amplitude on a specific size f law.  If a transducer requires more gain than the reference transducer, you know it is then a suspect.  This transducer can be used, providing you can still find the required defect called by the procedure. You  should  never  have  only  one  transducer  for  a  specific  application.    A  transducer  element  can  separate from the damping material.  This will cause the initial pulse to become a long ringing signal.  Such a situation will cause the transducer to fail the dead zone test.   The transducer must be replaced.  This type of process control will help ensure quality inspections without assuming everything is working properly.   Frequency of process control checks on equipment should come from the operations manual on the equipment.  Frequency of  transducer  checks  should  be  determined  by  the  amount  of  use,  which  will  be  determined  by  the  lab supervisor.  The operator is the critical link in this process.  Even if all the equipment is working properly, the inspector must follow the written procedure and use the correct standard.  No deviations SHALL be made without  proper  engineering  authority.    In  this  chapter,  the  terms  ‘‘reference  standard,’’  ‘‘reference  block,’’ ‘‘test  block,’’  and  ‘‘calibration  standard’’  all  have  the  same  meaning  as  defined  in  the  glossary.    Reference standards  are  used  by  the  instrument  operator.    Calibration  of  reference  standards  by  laboratories  is  not required.    However,  to  insure  uniform  inspection  sensitivity,  reference  standards  shall  be  traceable  to  a ‘‘master standard’’ in terms of discontinuity response. 1.8.1.1 Required Use. All  inspections  shall  include  the  use  of  one  or  more  reference  standards  for  setting  up  the  inspection.    In addition,  all  discontinuity  indications  shall  be  compared  to  a  reference  standard  by  comparing  the  signal amplitude of the discontinuity with the signal amplitude of the reference standard.  This is done either in percent signal amplitude or by noting the difference in amplitude in decibels (dB) when the instrument is equipped with dB attenuation controls. 1.8.1.2 Configuration. The reference standard may be a block containing a known size f lat-bottom or side-drilled hole, machined slot or notch, or a real discontinuity of known size in the test part or a piece similar to the test part.  Inspection procedures must be carefully reviewed for the following specific requirements: a. Flat surface reference standards used for test set-up and for evaluation of discontinuity size and metal travel  shall  be  fabricated  and  checked  in  accordance  with  ASTM  standard  practices.    For  more information see ASTM E127, E128, and E1158, 2nd para. 4.2.4.5. b. Curved  surface  reference  standards  may  be  required  when  performing  straight  beam  inspection  of curved  entry  surfaces  on  cylindrical  or  irregularly  shaped  products.    Special  ultrasonic  test  blocks containing specified radii of curvature and f lat-bottom holes of standard diameter shall be used.  For parts  with  convex  radii  over  4  inches,  use  standard  f lat  face  blocks.    Flat  blocks  may  be  used  to inspect other curved surfaces when supported by test data showing correction factors, and must be acceptable to the responsible engineering activity. 1-52 Change 2

Privacy Statement - Press Release - Copyright Information. - Contact Us - Support Integrated Publishing