SECTION VIII UT PROCESS CONTROL

T.O. 33B-1-1 c. Move the reference standard toward the magnet until a half-scale meter reading is obtained.  (If the south pole of the magnet is up, the meter needle will def lect to the right.) d. Mark this position on the sheet of paper by tracing along the bottom with a marker. e. Place  an  in-use  field  indicator  in  the  exact  position  traced  for  the  reference  standard  and  note  the reading. f. Reject an in-use indicator if the meter reading obtained differs from that of the reference standard by more than two units in either direction. Figure 1-9b. Checking In-Use Field Indicators SECTION VII - Deleted Change 3 1-51