Cutaway view of miniature oxygen regulator
Up
Aviation Structural Mechanic E2 - How airplanes are built and how to maintain them
Next
MD regulator operational drawing

Aircraft-Mounted  Oxygen  Regulators The  MD  series  regulator  is  being  used  in several multiplace naval aircraft. There are two types of regulators in this series—the MD-1 (low- pressure)  and  MD-2  (high-pressure).  The  only difference found in these regulators is operating pressure.  The  operating  pressure  of  the  MD-1 regulator is 50 to 500 psi. The pressure gauge reads 0 to 500 psi. The operating pressure of the MD-2 regulator is 50 to 2,000 psi. The pressure gauge reads 0 to 2,000 psi. The  following  controls  and  indicators  are located   on   the   front   panel   of   the   regulator (fig. 4-13). The small oblong-shaped window area on  the  left  side  of  the  panel  marked  FLOW indicates the flow of oxygen through the regulator by a visible blinking action. The pressure gauge is  found  on  the  upper  right  of  the  panel  and indicates  inlet  pressure  to  the  regulator.  The regulator has three control levers. A  supply  valve controller lever, located on the lower right corner, is used to control the supply of oxygen to the regulator; a diluter control lever, located on the lower center of the panel, has two positions— 100% OXYGEN and NORMAL OXYGEN; an emergency  pressure  control  lever, located on the lower  left  of  the  panel,  has  three  positions— EMERGENCY, NORMAL, and TEST MASK, and with the deluter lever in the 100% OXYGEN position, the regulator delivers 100 percent oxygen upon  inhalation  by  the  user.  In  the  NORMAL OXYGEN   position,   the   regulator   delivers   a mixture of air and oxygen with the air content decreasing until a cabin altitude of approximately 30,000   feet   is   reached.   Above   this   altitude, 100-percent  oxygen  is  delivered  to  the  user  upon inhalation. With  the  emergency  pressure  control  lever  in the  EMERGENCY  position,  the  regulator  delivers positive oxygen pressure to the outlet at altitudes when   positive   pressure   is   not   automatically delivered. In the TEST MASK position, oxygen is delivered to the mask under pressure too high to breathe and is used for checking the fit of the mask.  The  switch  must  be  in  the  NORMAL position to assure normal system operation. Refer to figure 4-14 for the operation of an MD type regulator. 1. Supply oxygen entering through the oxygen inlet (1) is filtered and passes through the manifold Figure 4-13.—Regulator operating controls and indicators. 4-18