![]() ![]() TM 11-6625-1702-35
c. Procedures.
Control
settings
Step
Equipment under test
No.
Test
equipment
Test procedures
Performance
standard
1 a. Turn on power a. Set METER FUNC- a. Adjust power supply to a. Test set meter indicates
TION
switch
to
obtain 27.5-volt indica-
27.5 ±1.3 volts dc.
supply,
and
28VDC.
tion on power supply
allow time for
meter.
warm up.
b. Set multimeter b. S e t METER FUNC- b. Adjust MV ADJUST b. Test set meter indicates
to 1-volt dc
TION switch to 0-500
for 250-mv indication
250 ±12 mv.
on multimeter.
range.
MV.
2 Set multimeter to Set METER FUNCTION Adjust MV ADJUST for Test set meter indicates
0.6-volt
dc
switch to 0-500 MV.
250-mv indication on
250 ±12.
multimeter.
range.
Figure 4-3. TUNE INDICATOR function test, test setup.
4-4
|
|